Skip to the main content
  • Fast delivery
  • Competitive prices
  • Made in The Netherlands

AOIP CALYS 50 Multifunction field calibrator

Datasheet View Catalog

Specifications

CALYS 50 is a basic multifunction calibrator within CALYS range. It is the perfect tool for advanced process maintenance and use on test bench in all industries. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process and frequency in one single instrument.
It provides extended functionalities (temperature simulation, scaling, steps, synthesizer, statistical functions…), easily accessible through quick menus by function.
The instrument simultaneously measures and simulates:Temperature: Up to 0,014% RDG
Resistance: Up to 0,012% RDG, 4 KΩ range
Current: Up to 0,0175% RDG, 50 mA range + 24 V loop supply
Voltage: Up to 0,013% RDG, 50 V IN / 20 V OUT range
Frequency: Up to 0,005% RDG, 20 KHz range (10 KHz in simulation)
RTD sensorsPt50, Pt100, Pt200, Pt500, Pt1000, Ni100, Ni120, Ni1000, Cu10 or Cu50
ThermocouplesK, T. J, E, R, S, U, B, L, C and N
DC current0-50mA
DC voltage100mV-50V
Frequency20kHz
Resistance400Ohm, resolution 1 mOhm
400Ohm, resolution 10 mOhm
DC current emission0-24mA
DC voltage emission100mV-20V
Frequency emission1000Hz-10kHz
Resistance emission40-4000Ohm
ConnectionConnect your probes by simply pushing on the terminal top and insert wires of up to 3 mm or 10 AWG diameter and compensated thermocouple connectors.
Wires are held tight between two brass plates ensuring thermal stability and a very good cold junction compensation for thermocouples.
This system also enables 4 mm banana plugs and security connectors to be connected on the terminal top.
Size210 x 110 x 50 mm
Weight900 g
Display240 x 320 pixel liquid crystal graphical display with backlite and contrast control Display of result as table of values or trend curve
Power supply230 V ± 10 %, 50/60 Hz
BatteryType: Lithium-ion, Charging Time: 3 hours, Lifetime: 8 hours
Communication portsUSB
Reference range23°C ± 5°C (RH: 45 to 75 % w/o condensing)
Operating reference range-10 to 50°C (RH: 20 to 80 % w/o condensing)
Limit operating range-15°C to +55°C (RH: 10 to 80 % w/o condensing) (70% at 55°C)
Storage temperature limits-30°C to +60°C
Maximum height0 to 2000 m
IP protectionIP54 according to EN60529
User manual
Battery charger
Set of 6 testing leads
Carrying strap
Factory test report

Customisation possibilities

You can remark the customisation in your quote