Plate Thermometer


Design
Application Fire testing
Requirements According to ISO 834 EN 1363 Part 1
Material Special treated heat resistant Plate
Dimensions 100 x 100 x 10 mm
Center part Specially treated for optimal cosine sensitivity
Connection 1/4 BSP female
Sheath
Construction Mineral insulated
Material Inconel 600
Insulation MgO
Diameter "D" 1 mm
Length "L" 2000 mm
Lead wire
PTFE insulated
flexible thermocouple wire,
braiding 260 °C
Wire length "K" 1000 mm
Element
Calibration Type K thermocouple
Accuracy According IEC 584-1 / DIN 43710
Thermocpuple Pre-aged
Hot junction Insulated
Testing Tested at 500 Volt/20 °C
Insulation resistance Minimum 100 M ohm
Ordering code  
Model   PL1001
   


 
Terug naar boven Terug Pagina emailen Download PDF zoeken Sitemap Inkoopvoorwaarden